Fenwal Inc: Amicus Exchange Kit Therapeutics. Component of Amicus Separator multiprocedural apheresis platform

Recall number
Z-1255-2023
Recalling firm
Fenwal Inc
Product
Amicus Exchange Kit Therapeutics. Component of Amicus Separator multiprocedural apheresis platform
Status
Open, Classified
Initiated
2023-02-14
Posted
2023-03-15
Terminated
not on file
Reason
Potential for centrifuge packs to develop a stress leak for certain lots of Amicus MNC Apheresis kit and Amicus Exchange kit (Therapeutic Kits) on the Amicus Separator.
Root cause
Process control
Action
Fresenius Kabi notified customers on 02/14/2023 via FedEx overnight delivery. Customers were instructed to evaluate affected inventory on site for potential continued use according to the conditions in the letter, inform potential users of affected product within the facility and any customers if further distributed, complete and return the acknowledgment form, and report any issues.
Quantity
6,564 units
Distribution
Nationwide distribution. International distribution to France, Korea, Taiwan, Colombia, India, Brazil, Germany, Canada.
Lot, serial or model codes
Product Code X6R2349; UDI: 04086000101950; Batch Numbers: FA22E23138, FA22F28168, FA22G28018, FA22H22225, FA22J28062, FA22K21065. The Lot number coding system is the following: FAXXYaabbc; FA = Haina (manufacturing / production unit); XX = two digit year; Y= letter of the alphabet corresponding to the month; aa= day; bb= sequential number; c= check digit.
510(k) numbers
["K180615"]
PMA numbers
not on file
Product code
LKN
Firm FEI
3004548776
Firm city
Lake Zurich
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register