FDA Medical Device Recalls (openFDA)
Randox Laboratories Ltd.: G-6-PDH Normal Control (G-6-PDH CONTROL N), Catalogue Number PD2618 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Glucose-6-Phosphate Dehydrogenase in erythrocytes
- Recall number
- Z-1255-2019
- Recalling firm
- Randox Laboratories Ltd.
- Product
- G-6-PDH Normal Control (G-6-PDH CONTROL N), Catalogue Number PD2618 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Glucose-6-Phosphate Dehydrogenase in erythrocytes
- Status
- Terminated
- Initiated
- 2019-04-03
- Posted
- not on file
- Terminated
- 2019-10-09
- Reason
- The target values and ranges in these lots are incorrect.
- Root cause
- Nonconforming Material/Component
- Action
- Randox sent an Urgent Medical Device Correction notification letters dated 4/2/19 to affected customers. The letter identified the affected product, problem and action to be taken: " Discontinue use of and quarantine any of the above devices immediately. " Discuss the contents of this notice with your Medical Director. " Review results generated with the affected batches in line with the clinical profile of the patient. " Inform all relevant staff members. If you have supplied or transferred any potentially affected product to another facility or organization, let that facility know of the recall immediately by providing a copy of this FSN and response form. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. For questions contact Randox Technical Services.
- Quantity
- 10 kits
- Distribution
- The products were distributed to the following US states: IL, OH, NC, PA, WI, WV and Puerto Rico.
- Lot, serial or model codes
- Lot Numbers: 676PD 701PD 716PD
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JBL
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28