Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: G-6-PDH Deficient Control (G-6-PDH CONTROL D), Catalogue Number PD2617 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Glucose-6-Phosphate Dehydrogenase in erythrocytes

Recall number
Z-1254-2019
Recalling firm
Randox Laboratories Ltd.
Product
G-6-PDH Deficient Control (G-6-PDH CONTROL D), Catalogue Number PD2617 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Glucose-6-Phosphate Dehydrogenase in erythrocytes
Status
Terminated
Initiated
2019-04-03
Posted
not on file
Terminated
2019-10-09
Reason
The target values and ranges in these lots are incorrect.
Root cause
Nonconforming Material/Component
Action
Randox sent an Urgent Medical Device Correction notification letters dated 4/2/19 to affected customers. The letter identified the affected product, problem and action to be taken: " Discontinue use of and quarantine any of the above devices immediately. " Discuss the contents of this notice with your Medical Director. " Review results generated with the affected batches in line with the clinical profile of the patient. " Inform all relevant staff members. If you have supplied or transferred any potentially affected product to another facility or organization, let that facility know of the recall immediately by providing a copy of this FSN and response form. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. For questions contact Randox Technical Services.
Quantity
14 kits
Distribution
The products were distributed to the following US states: IL, OH, NC, PA, WI, WV and Puerto Rico.
Lot, serial or model codes
Lot Numbers: 687PD 700PD 715PD
510(k) numbers
not on file
PMA numbers
not on file
Product code
JBL
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register