Nicolet Biomedical Div of Viasys Healthcare: VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes
- Recall number
- Z-1254-06
- Recalling firm
- Nicolet Biomedical Div of Viasys Healthcare
- Product
- VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes
- Status
- Terminated
- Initiated
- 2006-04-28
- Posted
- 2006-07-25
- Terminated
- 2006-12-27
- Reason
- The Bar Electrodes do not meet performance specification for electrode impedance. A manufacturing defect was discovered in disposable bar electrodes, part number 019-435600, lot 604664, that has the potential to reduce the amplitude of a motor or sensory nerve action potential. These possible lower amplitudes will be measured only if the bar electrode is used as a bar.
- Root cause
- Other
- Action
- Communication on April 28, 2006 by phone, followed by Certified Mail, with return receipt to confirm receipt of information . The letter describes the issue and asks that affected product be returned. The customer notification allows the user the option of using the Bar Electrode as 2 separate electrodes, as described in the Instructions for Use.
- Quantity
- 53 boxes, 10 electrodes per pouch, 10 pouches per box
- Distribution
- World wide including the states of AL, CA, CT, FL, IA, NH, NY, OH, TX. and countries of Canada, Japan, Poland
- Lot, serial or model codes
- Lot 604664
- 510(k) numbers
- ["K010019"]
- PMA numbers
- not on file
- Product code
- GXZ
- Firm FEI
- 3010611950
- Firm city
- Madison
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28