Skip to the content

FDA Medical Device Recalls (openFDA)

Nicolet Biomedical Div of Viasys Healthcare: VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes

Recall number
Z-1254-06
Recalling firm
Nicolet Biomedical Div of Viasys Healthcare
Product
VIASYS Healthcare Disposable Bar Electrodes. Re-order No. 019-435600. Single Use only. Manufactured in the U.S.A. for VIASYS Healthcare, 5225 Verona Road, Bldg 2, Madison, WI USA 53711-4495..Each box contains 10 pouches of Bar Electrodes and each pouch contains 10 electrodes
Status
Terminated
Initiated
2006-04-28
Posted
2006-07-25
Terminated
2006-12-27
Reason
The Bar Electrodes do not meet performance specification for electrode impedance. A manufacturing defect was discovered in disposable bar electrodes, part number 019-435600, lot 604664, that has the potential to reduce the amplitude of a motor or sensory nerve action potential. These possible lower amplitudes will be measured only if the bar electrode is used as a bar.
Root cause
Other
Action
Communication on April 28, 2006 by phone, followed by Certified Mail, with return receipt to confirm receipt of information . The letter describes the issue and asks that affected product be returned. The customer notification allows the user the option of using the Bar Electrode as 2 separate electrodes, as described in the Instructions for Use.
Quantity
53 boxes, 10 electrodes per pouch, 10 pouches per box
Distribution
World wide including the states of AL, CA, CT, FL, IA, NH, NY, OH, TX. and countries of Canada, Japan, Poland
Lot, serial or model codes
Lot 604664
510(k) numbers
["K010019"]
PMA numbers
not on file
Product code
GXZ
Firm FEI
3010611950
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register