Micro Therapeutics, Inc.: The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole catheter designed for the infusion of the Onyx liquid embolic system (LES). The catheter has a semi-rigid proximal shaft and a highly flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories. The outer diameter surface of the catheter is hydrophilically coated to increase lubricity. The AOMC is designed to facilitate catheter retrieval in the event the catheter becomes entrapped within the vasculature. The distal section of the catheter incorporates a detachment zone that allows detachment of the distal tip when the force required to extract the catheter exceeds the force to detach the tip. The catheter has two radiopaque marker bands to visualize the position of the catheter and the detach zone area, one proximal to the detach zone and one at the distal end of the catheter.

Recall number
Z-1253-2024
Recalling firm
Micro Therapeutics, Inc.
Product
The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole catheter designed for the infusion of the Onyx liquid embolic system (LES). The catheter has a semi-rigid proximal shaft and a highly flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories. The outer diameter surface of the catheter is hydrophilically coated to increase lubricity. The AOMC is designed to facilitate catheter retrieval in the event the catheter becomes entrapped within the vasculature. The distal section of the catheter incorporates a detachment zone that allows detachment of the distal tip when the force required to extract the catheter exceeds the force to detach the tip. The catheter has two radiopaque marker bands to visualize the position of the catheter and the detach zone area, one proximal to the detach zone and one at the distal end of the catheter.
Status
Open, Classified
Initiated
2024-01-11
Posted
2024-03-04
Terminated
not on file
Reason
European version of microcatheter were distributed within US which contain a different "Indications for Use".
Root cause
Under Investigation by firm
Action
On January 11, 2024, Medtronic Neurovascular issued a "Urgent: Medical Device Recall" Medtronic ask consignees to take the following action: 1. Do not use any impacted product. Remove and quarantine all unused impacted product in your inventory. 2. Return the impacted products to Medtronic. Your Medtronic representative may assist in facilitating the return of product, as necessary. If an alternative product is needed, your Medtronic representative can assist you with identifying a suitable replacement product. 3. Complete and return the Customer Confirmation form enclosed with this letter acknowledging that you have received this information.
Quantity
95 units
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, MA, MD, MI, MO, NC, ND, NE, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI.
Lot, serial or model codes
UDI: 00763000311445/ Lot # B510758, B555761, B615024, B615025, B629693, B629694
510(k) numbers
not on file
PMA numbers
not on file
Product code
MFE
Firm FEI
1000220862
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register