Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models 11061620 & 11061628) SOMATOM go.All (Model 11061630) SOMATOM go.Top (Models 11061640 & 11061648) SOMATOM go.Sim (Model 11061660) SOMATOM go.Open Pro (Model 11061670) SOMATOM X.cite (Model 11330001)

Recall number
Z-1253-2021
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models 11061620 & 11061628) SOMATOM go.All (Model 11061630) SOMATOM go.Top (Models 11061640 & 11061648) SOMATOM go.Sim (Model 11061660) SOMATOM go.Open Pro (Model 11061670) SOMATOM X.cite (Model 11330001)
Status
Terminated
Initiated
2021-03-03
Posted
not on file
Terminated
2023-12-28
Reason
SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in diagnosis or patient rescan may occur. Sporadic software errors during interventional workflows may also result
Root cause
Software design
Action
Siemens Medical Solutions USA, Inc., provided Customer Information Letter on 3/3/21 to affected customers via CT063/20/S. Customers with valid email addresses emailed a copy of the CSAN via Adobe Mega Sign. The CSAN will be distributed to the remaining customers address via hand delivery by the Siemens Customer Service Engineer. The letter states reason for recall, health risk and action to take: Siemens Healthineers has developed software update syngo.CT VA30A_SP3 to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also provide improvements related to the general system, examination and reconstruction processes, as well as improvements for dual energy post-processing. Following the corrective action, the cause has been eliminated and recurrence of the identified issues are prevented. This software update will be provided to you free of charge. How will the corrective action be implemented? Software update CT084/20/S will be performed onsite for SOMATOM X.cite (Model #11330001). For all other systems, the update CT064/20/S will be performed remotely. The remote software updates process will require approximately 180 minutes for completion. If you have any unresolved questions or you require technical support, please contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Quantity
190 units US
Distribution
US Nationwide distribution.
Lot, serial or model codes
Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a
510(k) numbers
["K163296","K173632","K192061","K200524"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register