Siemens Healthcare Diagnostics Inc: RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857, 10697306. The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili.
- Recall number
- Z-1253-2015
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Product
- RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857, 10697306. The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili.
- Status
- Terminated
- Initiated
- 2014-12-30
- Posted
- 2015-03-06
- Terminated
- 2017-08-02
- Reason
- Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
- Root cause
- Software design
- Action
- An Urgent Field Safety Notice was sent by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on December 30, 2014 both in the United States and elsewhere for communication with affected customers. In addition, all affected domestic customers were sent a hard copy of the Urgent Field Safety Notice via Federal Express on December 30, 2014. All affected customers outside the US will be provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs consignees of the issue. Consignees are also asked to complete a field effectiveness check form and return it Siemens via fax (312) 275-7795.
- Quantity
- 4376 devices
- Distribution
- Worldwide Distribution - US nationwide including Puerto Rico, Canada, Mexico, Albania, Algeria, Argentina, Austria, Bahrain, Belgium, Bosnia Herzegovina, Botswana, Brazil, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iraq, Ireland, Israel, Italy, Ivory Coast, Kazakhstan, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Netherlands, New Caledonia, Norway, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, U.A.E., United Kingdom, Vatican City, Venezuela, White Russia/Belarus.
- Lot, serial or model codes
- All serial numbers.
- 510(k) numbers
- ["K113216"]
- PMA numbers
- not on file
- Product code
- MQM
- Firm FEI
- 3002637618
- Firm city
- Norwood
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28