Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Renal Div: Ultra-Filtration (UF) Removal Regulators(Rgltr Assy UF REM Molded Spare); a spare part used with Baxter''s System 1000, Tina, Arena, Altratouch and Aurora Hemodialysis Instruments; Baxter Healthcare Corporation, Deerfield, IL 60015; item numbers 6001276024 and 6001276028

Recall number
Z-1253-06
Recalling firm
Baxter Healthcare Renal Div
Product
Ultra-Filtration (UF) Removal Regulators(Rgltr Assy UF REM Molded Spare); a spare part used with Baxter''s System 1000, Tina, Arena, Altratouch and Aurora Hemodialysis Instruments; Baxter Healthcare Corporation, Deerfield, IL 60015; item numbers 6001276024 and 6001276028
Status
Terminated
Initiated
2006-06-21
Posted
2006-07-25
Terminated
2008-03-04
Reason
The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 06/21/06 to all System 1000, TINA, Altratouch, Arena and Aurora Hemodialysis Instrument customers who may have received the nonconforming assemblies via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the three lots of UF Removal Regulators were potentially assembled incorrectly. While these regulators are identical in appearance to properly assembled components, the mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy. Replacement regulators were included with the recall letter, and the accounts were requested to return the regulators from their previous shipment to Baxter using the enclosed mailing materials. The accounts were requested to complete the enclosed reply form indicating the number of units received and the number of units returned and acknowledging receipt of the letter and dissemination of the information to their staff and to other services of facilities, as applicable, and fax the sheet to 1-727-544-3025. Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 1.
Quantity
569 units
Distribution
Worldwide Distribution including the states of Arizona, California, Connecticut, Florida, Georgia, Maryland, Missouri, Ohio, Oklahoma, Tennessee, Texas and Puerto Rico, and international distribution to Belgium, Canada, Thailand, Chile, Austrailia, Peru, Argentina, Panama and Mexico.
Lot, serial or model codes
item number 6001276024, lot numbers 01095126 and 01092479; part number 6001276028, lot number 01095127
510(k) numbers
["K030099","K103562","K910215","K970446"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.