FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1252-2011
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Status
- Terminated
- Initiated
- 2011-01-05
- Posted
- 2011-02-15
- Terminated
- 2011-04-07
- Reason
- Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.
- Root cause
- Other
- Action
- The firm, Elekta Limited, sent an "Important Notice A344" dated January 5, 2011 to all consignees/customers. The notice described the product, problem and action to be taken by the customers. The customers were instructed to follow any instructions or recommendations covered in the notice. The notice provides suggestions on how to check and prevent this situation from occurring; it instructs the User to file the document in the Important Notice section of the appropriate User Manual; to complete a FlexMap Check each time they acquire a new FlexMap and amend their local procedures to include the FlexMap Check and complete and return the Important Notice Confirmation of Receipt via Fax to: Field Support Administration, Elekta Limited, Linac House, Fleming Way Crawley, West Sussex, RH10 9RR UK +44 (0)1293 654401. If you have any questions, contact Elekta Limited at +44 (0)1293 654200 or www.elekta.com.
- Quantity
- 223 units
- Distribution
- Nationwide distribution.
- Lot, serial or model codes
- 105257, 105803, 105941, 105982, 105984, 151024, 151038, 151051, 151053, 151055, 151073, 151074, 151084, 151110, 151130, 151141, 151151, 151156, 151157, 151160, 151168, 151173, 151178, 151216, 151245, 151250, 151256, 151258, 151259, 151260, 151298, 151301, 151309, 151317, 151319, 151326, 151327, 151328, 151340, 151352, 151359, 151360, 151362, 151363, 151371, 151387, 151391, 151394, 151396, 151398, 151410, 151411, 151412, 151416, 151418, 151423, 151432, 151433, 151435, 151441, 151443, 151446, 151467, 151479, 151504, 151507, 151514, 151515, 151517, 151519, 151521, 151522, 151523, 151530, 151538, 151540, 151545, 151557, 151558, 151566, 151567, 151568, 151574, 151584, 151585, 151587, 151595, 151598, 151601, 151611, 151616, 151623, 151626, 151627, 151628, 151632, 151634, 151636, 151640, 151653, 151664, 151672, 151674, 151675, 151678, 151683, 151684, 151689, 151694, 151695, 151696, 151698, 151699, 151700, 151704, 151705, 151708, 151710, 151714, 151722, 151724, 151728, 151729, 151734, 151736, 151751, 151759, 151760, 151763, 151764, 151765, 151770, 151774, 151779, 151782, 151783, 151785, 151789, 151791, 151793, 151801, 151802, 151803, 151805, 151808, 151809, 151810, 151811, 151812, 151820, 151828, 151839, 151842, 151847, 151848, 151855, 151860, 151864, 151866, 151867, 151875, 151881, 151882, 151885, 151886, 151892, 151931, 151933, 151942, 151952, 151953, 151955, 151956, 151958, 151959, 151960, 151961, 151965, 151969, 151977, 151978, 151981, 151989, 152008, 152019, 152023, 152031, 152043, 152051, 152064, 152079, 152080, 152091, 152098, 152103, 152115, 152116, 152117, 152123, 152126, 152154, 152158, 152163, 152174, 152175, 152176, 152177, 152192, 152200, 152210, 152211, 152214, 152217, 152220, 152222, 152223, 152232, 152246, 152257, 152271, 152301, 152307, 152308
- 510(k) numbers
- ["K051932"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28