Skip to the content

FDA Medical Device Recalls (openFDA)

MAQUET Inc.: Servo Ventilator 300 (Model SV300) and Servo Ventilator 300 (Model SV300A). The SV300 is intended for general and critical ventilatory care for use on adults, pediatric, infant and neonatal patients. The unit is designed to be used at the bedside for intra-hospital transport.

Recall number
Z-1252-2009
Recalling firm
MAQUET Inc.
Product
Servo Ventilator 300 (Model SV300) and Servo Ventilator 300 (Model SV300A). The SV300 is intended for general and critical ventilatory care for use on adults, pediatric, infant and neonatal patients. The unit is designed to be used at the bedside for intra-hospital transport.
Status
Terminated
Initiated
2009-03-23
Posted
2009-05-12
Terminated
2009-11-16
Reason
The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails.
Root cause
Device Design
Action
An "Urgent: Device Correction" letter dated March 23, 2009 was sent by Federal Express to customers. The letter described the affected product, problem, indications, potential hazard and actions for users. Customers were instructed to complete the "Device Correction -- Customer Fax Response" form and return via fax (732-667-1903) to Maquet Inc. RA/QA. Direct questions to your local Maquet Inc. representative.
Quantity
7,847 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
All units manufactured prior to June 2002.
510(k) numbers
["K902859","K970839"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3008355164
Firm city
Bridgewater
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register