Skip to the content

FDA Medical Device Recalls (openFDA)

NeuMoDx Molecular Inc: NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900

Recall number
Z-1251-2022
Recalling firm
NeuMoDx Molecular Inc
Product
NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900
Status
Open, Classified
Initiated
2022-05-13
Posted
not on file
Terminated
not on file
Reason
There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment.
Root cause
Process control
Action
On May 13, 2022, the firm distributed Urgent Medical Device Correction letters to impacted customers. The letter informed customers of the reduced analytical sensitivity for Flu A, SARS-CoV-2, and RSV. The firm also specified that only the RSV target showed a similar reduction in clinical sensitivity; Influenza A and SARS-CoV-2 did not show a reduction in clinical sensitivity. Influenza B is not affected by this issue. The letter reminded users that negative results do not preclude infection with Influenza virus, RSV, and/or SARS-CoV-2, and the device should not be used as the sole basis for patient management and treatment decisions. Customers should share the notice with all affected users. If you have any questions regarding this recall, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com.
Quantity
459 kits US; 1626 kits OUS
Distribution
Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI.
Lot, serial or model codes
UDI-DI (GTIN): 10814278020830 Cartridge lot numbers 111915 - 117508
510(k) numbers
not on file
PMA numbers
not on file
Product code
QLT
Firm FEI
3010729479
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register