FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Modular Replacement System (MRS) Cemented Stems, Sterile, Catalog Numbers: 6485-3-000, 6485-3-008, 6485-3-009, 6485-3-010, 6485-3-018, 6485-3-019, 6485-3-300, 6485-3-308, 6485-3-309, 6485-3-310, 6485-3-318, and 6485-3-319. The Modular Replacement System has been successfully used in clinical in clinical situations where there is extensive bone loss due to tumor resection, and/or failed previous prosthesis, or trauma. These components can be used to replace the proximal femur, the distal femur, the proximal tibia, or to reconstruct the total femur in extreme clinical situations.
- Recall number
- Z-1251-2009
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Modular Replacement System (MRS) Cemented Stems, Sterile, Catalog Numbers: 6485-3-000, 6485-3-008, 6485-3-009, 6485-3-010, 6485-3-018, 6485-3-019, 6485-3-300, 6485-3-308, 6485-3-309, 6485-3-310, 6485-3-318, and 6485-3-319. The Modular Replacement System has been successfully used in clinical in clinical situations where there is extensive bone loss due to tumor resection, and/or failed previous prosthesis, or trauma. These components can be used to replace the proximal femur, the distal femur, the proximal tibia, or to reconstruct the total femur in extreme clinical situations.
- Status
- Terminated
- Initiated
- 2009-01-30
- Posted
- 2009-04-17
- Terminated
- 2010-04-08
- Reason
- Instruction for use incorrect: There are no statements or warning on labels or Instructions for Use that 8, 9 and 10 mm diameter MRS Cemented Stems are indicated only for use in the GMRS Proximal Tibial and Distal Femoral Components.
- Root cause
- Labeling False and Misleading
- Action
- Urgent Product Recall letters, dated 1/30/2009, were sent to all Branches/agencies, Hospital Risk Management, Chief of Orthopaedics and Surgeons on January 30, 2009 via Federal Express. The letter asked consignees to examine their inventory and determine if they had any affected products, remove and quarantine affected product from their inventory, and complete the attached Product Recall Acknowledgment Form. The letter also stated that a Stryker Orthopaedics Sales representative would contact the consignee to reconcile and return the product . Consignees should contact their Stryker sales rep if they weren't not contacted by them. Questions should be directed to Colleen O'Meara at 201-831-5970.
- Quantity
- 1161 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- All lots
- 510(k) numbers
- ["K040749"]
- PMA numbers
- not on file
- Product code
- KRO
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28