FDA Medical Device Recalls (openFDA)
Instrumentation Laboratory Co.: Hemoliance RecombiPlasTin. Prothrombin Time Test. The product is available as a kit of 25 vials of Hemoliance RecombiPlasTin matched with 25 vials of Hemoliance RecombiPlasTin Diluent. Hemoliance RecombiPlasTin after reconstitution with Hemoliance RecombiPlasTin Diluent is a liposomal preparation of recombinant human tissue factor and purified phospholipids, calcium chloride, buffer and a preservative. The use of recombinant human tissue factor and purified phospholipids assures uniformity in reagent purity, ISI values and performance from lot to lot. Once reconstituted, it is ready for use in the one-stage PT test. Hemoliance RecombiPlasTin Diluent is an aqueous solution of calcium chloride and a preservative. --- FOR IN VITRO DIAGNOSTIC USE.
- Recall number
- Z-1247-06
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- Hemoliance RecombiPlasTin. Prothrombin Time Test. The product is available as a kit of 25 vials of Hemoliance RecombiPlasTin matched with 25 vials of Hemoliance RecombiPlasTin Diluent. Hemoliance RecombiPlasTin after reconstitution with Hemoliance RecombiPlasTin Diluent is a liposomal preparation of recombinant human tissue factor and purified phospholipids, calcium chloride, buffer and a preservative. The use of recombinant human tissue factor and purified phospholipids assures uniformity in reagent purity, ISI values and performance from lot to lot. Once reconstituted, it is ready for use in the one-stage PT test. Hemoliance RecombiPlasTin Diluent is an aqueous solution of calcium chloride and a preservative. --- FOR IN VITRO DIAGNOSTIC USE.
- Status
- Terminated
- Initiated
- 2006-03-07
- Posted
- 2006-07-18
- Terminated
- 2006-08-11
- Reason
- Firm received 2 complaints concerning incorrect prothrombin time (PT) results while using the reagents on samples of patients taking the antibiotic CUBICIN (Daptomycin for injection).
- Root cause
- Other
- Action
- "Urgent Product Notifications" were sent to IL customers on 3/07/06 through its distributor, Beckman Coulter.
- Quantity
- 5 mL: 3870; 20 mL: 29,007 (all Domestic)
- Distribution
- Nationwide
- Lot, serial or model codes
- Product No.49732750 (5 mL): Lots N0448239, N0549190, N0940581, N0353406, N0857200, N1050152; Product No. 49732720 (20 mL): Lots N0347895, N0548959, N0749495, N0940543, N0940586, N1242409, N0353410, N0454257, N0857199, N1050143.
- 510(k) numbers
- ["K925604"]
- PMA numbers
- not on file
- Product code
- GJS
- Firm FEI
- 1217183
- Firm city
- Lexington
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28