Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Recall number
Z-1246-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
Status
Terminated
Initiated
2019-02-14
Posted
not on file
Terminated
2020-05-01
Reason
Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.
Root cause
Software design
Action
Siemens issued letter dated 2/14/19 providing reason for recall and a user workaround in the Customer Safety Advisory Notice (CSAN) CT006/19/S. The recall issue will be permanently resolved by software update syngo CT VB10A_SP1. Contact the service organization 1-800-888-7436.
Quantity
53 total units distributed in U.S.
Distribution
US Nationwide Distribution
Lot, serial or model codes
Serial numbers of all affected models: 122022 100143 100107 83834 83746 83811 66649 122020 83801 122008 122029 83767 83852 100092 83798 100091 100112 100129 100136 83837 122023 83871 100117 100097 100099 122009 83763 83819 83752 83845 83846 83749 83847 100146 83786 83842 83775 100090 83796 83768 83859 100087 83826 83754 83790 122011 83824 100147 83810 83758 100140 100085 83808
510(k) numbers
["K173607"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register