Stryker Instruments Div. of Stryker Corporation: 1. InterPulse Handpiece with coaxial bone cleaning tip, Product Number: 0210-110-100, UDI: (01)4546540861641; 2. InterPulse Handpiece with Coaxial High Flow Tip, Product Number: 0210-114-100, UDI: (01)4546540861764; 3. InterPulse Handpiece with Retractable Coaxial Fan Spray Tip, Product Number: 0210-118-200, UDI: (01)4546540861818; 4. InterPulse Irrigation Only Handpiece with Fan Spray Tip, Product Number: 0210-218-000, UDI: (01)4546540144300 for orthopedic use.

Recall number
Z-1246-2018
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
1. InterPulse Handpiece with coaxial bone cleaning tip, Product Number: 0210-110-100, UDI: (01)4546540861641; 2. InterPulse Handpiece with Coaxial High Flow Tip, Product Number: 0210-114-100, UDI: (01)4546540861764; 3. InterPulse Handpiece with Retractable Coaxial Fan Spray Tip, Product Number: 0210-118-200, UDI: (01)4546540861818; 4. InterPulse Irrigation Only Handpiece with Fan Spray Tip, Product Number: 0210-218-000, UDI: (01)4546540144300 for orthopedic use.
Status
Terminated
Initiated
2017-11-09
Posted
not on file
Terminated
2019-04-16
Reason
Lack of sterility assurance as bioburden levels were found to be higher than internal acceptable rates.
Root cause
Environmental control
Action
Letters dated November 16, 2017, were sent to customers. Actions to be taken by the Customer/User: 1. Immediately review this Recall Notification. 2. Check all stock areas and/or operating room storage to determine if any devices from the affected lots are at your facility. 3. Quarantine and discontinue use of the recalled devices. 4. Complete and sign the enclosed Business Reply Form (BRF). 5. If you have further distributed these products, please forward this letter and the attached BRF to all affected locations. Please indicate each location on the BRF. 6. Return the completed BRF to Stryker Instruments Post Market Regulatory Compliance via fax (866-521- 2762) or email (erin.bissonnette@stryker.com). 7. If the BRF for your facility indicates that recalled product is currently on hand, a FedEx label will be emailed to you. This shipping label should be used to return recalled product. Upon receipt of the recalled product, a credit will be issued to your account. 8. Report any adverse events associated with the use of these devices to Stryker Instruments 1-800-253-3210, including any infections potentially associated with the use of affected product.
Quantity
5988
Distribution
Distributed in all 50 U.S. States and the District of Columbia. Distributed in Australia, Argentina, Japan, Canada, Chile, Korea, Spain, France, Italy, Netherlands, and United Kingdom.
Lot, serial or model codes
1. Lot Numbers: 17283012, 17284012, 17286012; 2. Lot Numbers: 17278012, 17280012, 17282012, 17283012, 17285012, 17289012, 17292012; 3. Lot Number: 17279012; 4. Lot Number: 17284012   UDI: (01)4546540861641; (01)4546540861764; (01)4546540861818; (01)4546540144300
510(k) numbers
not on file
PMA numbers
not on file
Product code
FQH
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register