Ethicon Endo-Surgery Inc: ENSEAL 5 mm Diameter Tissue Sealer G2 14 cm Length Straight Jaw, Model # NSLG2S14 and ENSEAL 5 mm Diameter Tissue Sealer G2 25 cm Length Straight Jaw, Model # NSLG2S25. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 The ENSEAL G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic, general and gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the instruments.
- Recall number
- Z-1246-2012
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- ENSEAL 5 mm Diameter Tissue Sealer G2 14 cm Length Straight Jaw, Model # NSLG2S14 and ENSEAL 5 mm Diameter Tissue Sealer G2 25 cm Length Straight Jaw, Model # NSLG2S25. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 The ENSEAL G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic, general and gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the instruments.
- Status
- Terminated
- Initiated
- 2012-01-26
- Posted
- 2012-03-21
- Terminated
- 2013-02-15
- Reason
- Ethicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation: The ENSEAL device/system may continue to activate when the activation button is released prior to reaching the end of cycle (tone 3). (ii) No Activation
- Root cause
- Nonconforming Material/Component
- Action
- Ethicon Endo-Surgery sent an Urgent Medical Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop use of the affected product, fill out the attached Business Reply Form (BRF) and return to and return the affected product to: Ethicon Endo-Surgery ATT: ENSEAL G2 Recall 4545 Greek Road Cincinnati, Ohio 45242 Customers were instructed to choose one of the following response options: Return the Business Reply Form to their sales representative Call 1-800-873-3636, Option 6 Fax the BRF to 1-513-337-4138 For any questions regarding this recall call 513-337-3419.
- Quantity
- 84 pieces
- Distribution
- Worldwide Distribution - USA including AZ, CA, CT, FL, IL, IN, KS, KY, MA, MI, MO, NC, NY, OH, PA, TN & TX and the countries of Arab Emirates, Austria, Belgium, France, Germany, Great Britain, Italy, Jordan, Kingdom of Saudi Arabia, Lebanon, Portugal, Slovenia, Sweden and Turkey.
- Lot, serial or model codes
- Product Code 00NSLG2S14, Lot #, H44Z5R, Exp Date, Nov-2013; Product Code 00NSLG2S25, Lot #, H44Z57, Exp Date, Nov-2013 & Lot # J4A28X, Exp Date, Dec-2013
- 510(k) numbers
- ["K112033"]
- PMA numbers
- not on file
- Product code
- LFL
- Firm FEI
- 1527736
- Firm city
- Cincinnati
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28