Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer TM-400 Broach Handle; Part No. 96-210-10001. The product is a surgical instrument for use in conjunction with Zimmer TM-400 Trabecular Metal implants during spinal surgery. The Broach Handle threads onto various broaches and trial implants. The handle with these broaches or trials attached is used to prepare and measure a vertebral space to accept a TM-400 Trabecular Metal implant.

Recall number
Z-1246-2009
Recalling firm
Zimmer Inc.
Product
Zimmer TM-400 Broach Handle; Part No. 96-210-10001. The product is a surgical instrument for use in conjunction with Zimmer TM-400 Trabecular Metal implants during spinal surgery. The Broach Handle threads onto various broaches and trial implants. The handle with these broaches or trials attached is used to prepare and measure a vertebral space to accept a TM-400 Trabecular Metal implant.
Status
Terminated
Initiated
2009-04-29
Posted
2009-07-07
Terminated
2010-10-08
Reason
The broach handle may fracture at the threaded tip during use, which would result in surgical intervention to retrieve the broach or the trial as the threaded tip fragment prevents attachment of another handle to retrieve the device.
Root cause
Device Design
Action
Distributors and user accounts were notified by "Urgent: Medical Device Recall" letter dated April 29, 2009. The letter described the affected product, reason for recall, health risks, and action for consignees. Distributors were instructed to retrieve the devices and to return them to Zimmer no later than May 13, 2009. This instrument recall will result in an inability to implant the associated devices until a new instrument becomes available. For questions or assistance about the recall please contact Zimmer Spine by calling 800-777-7505.
Quantity
435
Distribution
Nationwide, Austria, Australia, Belgium, Canada, Finland, Germany, Greece, Korea, South Africa, Sweden, Switzerland, Taiwan and the United Kingdom.
Lot, serial or model codes
All lots.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTQ
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28