Aesculap Inc: DISP.TROCAR W.DILATING PIN 12/110MM, Product Code EK234SU. For use in laparoscopic procedures.

Recall number
Z-1245-2024
Recalling firm
Aesculap Inc
Product
DISP.TROCAR W.DILATING PIN 12/110MM, Product Code EK234SU. For use in laparoscopic procedures.
Status
Open, Classified
Initiated
2024-01-09
Posted
2024-03-01
Terminated
not on file
Reason
The sterile blister packaging may be damaged, and sterility may be compromised.
Root cause
Packaging change control
Action
URGENT MEDICAL DEVICE RECALL NOTIFICATION letters dated January 5, 2024 were sent to customers. Customers were instructed to cease use and quarantine all product subject to recall. Once Aesculap Customer Support receives the customer's acknowledgement form, they will contact the customer to arrange for credit and to have the current inventory returned.
Quantity
6 pieces
Distribution
Worldwide - US Nationwide distribution in the states of KS, NJ and NY. The country of Canada.
Lot, serial or model codes
Product Code: EK234SU; UDI/DI: 04046963620462; Batch Numbers (Expiration Date): 52573073 (12/08/2023);
510(k) numbers
["K101937"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register