Skip to the content

FDA Medical Device Recalls (openFDA)

Tosoh Smd Inc: Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.

Recall number
Z-1245-2017
Recalling firm
Tosoh Smd Inc
Product
Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.
Status
Terminated
Initiated
2016-12-22
Posted
not on file
Terminated
2019-06-18
Reason
Power outages causes reporting software to shutdown.
Root cause
Software design
Action
Tosoh Bioscience sent a Customer Recall Notification letter dated December 21, 2016 to customers. The letter identified the affected product, problem and action to be taken. For questions call Tosoh Technical Support at 800-248-6764.
Quantity
34
Distribution
US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.
Lot, serial or model codes
All G8's with Reporting Software interfaced with LIS.
510(k) numbers
not on file
PMA numbers
not on file
Product code
PQQ
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register