Skip to the content

FDA Medical Device Recalls (openFDA)

Nidek Inc: RT-5100 (Epic 5100) Refractor and RT 3100 Refractor: Product Usage: Instrument providing means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subjects eye for purpose of determining refractive error and binocular functions.

Recall number
Z-1245-2016
Recalling firm
Nidek Inc
Product
RT-5100 (Epic 5100) Refractor and RT 3100 Refractor: Product Usage: Instrument providing means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subjects eye for purpose of determining refractive error and binocular functions.
Status
Terminated
Initiated
2016-02-29
Posted
2016-03-24
Terminated
2018-09-24
Reason
There has been a reoccurrence of an issue on the RT 5100 and RT 3100 Refractor of the Near Point Chart Arm lowering spontaneously with potential for slight injury to face or nose of patient.
Root cause
Component design/selection
Action
Nidek sent a Recall letters dated February 29, 2016 to all affected site. The letter identified the affected product, problem and actions to be taken. Nidek Customer Service will be replacing the affected product with new parts to address this recall. For questions call Customer Service at 1-800-722-0219.
Quantity
2703 separate serial numbers
Distribution
US Nationwide Distribution
Lot, serial or model codes
serial numbers: See attached distribution list. 128 RT -3100 and 2565 RT-5100.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HKN
Firm FEI
2936921
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register