Intuitive Surgical, Inc.: daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system.

Recall number
Z-1245-2007
Recalling firm
Intuitive Surgical, Inc.
Product
daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system.
Status
Terminated
Initiated
2007-05-14
Posted
2007-09-26
Terminated
2007-12-03
Reason
Product may malfunction and fail to start up on AC power.
Root cause
Process control
Action
As of 9/12/2007, the firm has conducted a field inspection of any involved units via service visit. No other notice is planned or indicated by the firm's submission. The firm feels that the risk associated with this issue is minimal, and therefore, any advanced notification is unneccessary. If any system failed one or more of the inspection points, its APB (3510404-04) was replaced with a verified APB (351040-04 or 351040-05).
Quantity
38 units
Distribution
Worldwide, including USA, France, China, Japan, Mexico, Spain, Italy, Belgium and Singapore.
Lot, serial or model codes
Model number IS2000, part number 351040-04. System numbers SG002, SG004, SG010 , SG032, SG060, SG061, SG064, SG065, SG066, SG067, SG068, SG069, SG070, SG071, SG072, SG073, SG074, SG075, SG076, SG077, SG078, SG079, SG080, SG081, SG082, SG083, SG084, SG086, SG090, SG116, SG117, SG118, SG120, SG121, SG126, SG174, SG175, SG184
510(k) numbers
["K050369","K063220"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register