Bard Peripheral Vascular Inc: Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239

Recall number
Z-1244-2025
Recalling firm
Bard Peripheral Vascular Inc
Product
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Status
Open, Classified
Initiated
2025-02-05
Posted
2025-03-05
Terminated
not on file
Reason
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.
Root cause
Under Investigation by firm
Action
On 2/5/2025, "URGENT: MEDICAL DEVICE CORRECTION" notices were sent to customers informing them of the following: 1) In order to minimize risks of illness or injury, please review the updates to the eIFU in their entirety. The updated eIFU can be found on BD's website at https://www.bardpv.com/Rotarex-IFU.php 2) Firm recommends that end users review the updated on-line, on-demand training on the safe and effective use of the Rotarex Atherectomy System, reflecting the recent eIFU updates. The online education may be completed by registering on the BDLA Learning Management System https://academy.bd.com. 3) If possible, post this notice with the stored product as evidence of the updated eIFU. 4) Share this notification with all users within your facility network of the product to ensure they are also aware of the updated eIFU. This correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 5) Complete and return the response form via email to BDRC20@bd.com Report any complaints experienced with the use of this product to the firm's Regional Complaint Center: 1-844-8BD-LIFE (1-844-823-5433) Say "Product Complaints" when prompted Mon-Fri 8:00am to 5:00pm CT Email: productcomplaints@bd.com For additional information regarding Advance Clinical Education and Training Programs, please email the Advance Team at bdpi.advance@bd.com Distributors were provided with a copy of the customer communication (correction notice) and asked to provide a copy to all customers on BD's behalf. If you have any questions, call 480-303-2602.
Quantity
15,755
Distribution
US Nationwide Distribution: NY, VA, KY, IN, OR, FL, MD, NJ, LA, CO, PR, PA, MS, MI, TX, KS, IL, CA, GA, OH, AL, NC, ND, AZ, SC, TN, UT, AR, WI, SD, MA, MO, ME, DE, WV , NV, NH, MN, CT, NE, IA, NM, WA, HI, and OK
Lot, serial or model codes
Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/230893(2025-12-15), 231100(2026-06-09), 231172(2026-06-09), 231171(2026-06-09), 231526(2026-08-21), 231875(2026-08-21), 240018(2026-11-22), 240017(2026-11-22), 240375(2027-01-18), 240567(2027-01-18), 241075(2024-05-22), 241335(2024-06-19), 241550(2024-07-19), 241777(2024-08-19), 242691(2024-12-13); 80237/07640142811862/230897(2026-04-11), 230898(2026-04-11), 231033(2026-05-31), 231101(2026-06-13), 231170(2026-06-13), 231229(2026-07-07), 231400(2026-07-14), 231414(2026-07-31), 231412(2026-06-30), 231413(2026-07-31), 231411(2026-08-03), 231521(2026-09-01), 231520(2026-07-31), 231522(2026-08-31), 231523(2026-09-01), 231524(2026-09-01), 231525(2026-09-01), 231715(2026-09-22), 231885(2026-09-22), 231884(2026-10-27), 240020(2026-11-22), 240396(2026-12-08), 240530(2027-01-10), 240531(2027-01-10), 240566(2027-01-10), 240568(2027-01-25), 240754(2027-02-06), 240769(2027-02-09), 240795(2027-03-04), 241074(2027-03-07), 241119(2027-03-21), 241190(2027-04-16), 241325(2027-04-16), 241328(2027-04-25), 241336(2027-04-29), 241464(2027-05-07), 241471(2027-05-27), 241553(2027-06-05), 242019(2027-06-28), 242023(2027-06-12), 242057(2027-06-20), 242056(2027-06-20), 242112(2027-06-26), 242160(2027-07-25), 242321(2027-07-08), 242493(2027-08-01), 242596(2027-09-20), 242693(2027-11-06); 80238/07640142811879/230895(2025-11-23), 240128(2026-05-31), 240276(2026-05-31), 240532(2026-05-31), 240658(2026-09-30), 241076(2026-09-30), 241330(2026-09-30), 241548(2026-09-30), 241768(2026-09-30), 241911(2026-09-30), 242117(2026-09-30), 242494(2026-09-30), 242521(2026-09-30), 242519(2026-05-31), 242520(2026-09-30); 80239/07640142811886/230896(2025-11-23), 231231(2026-06-13), 231410(2026-06-13), 231527(2026-08-31), 231883(2026-10-19), 232029(2026-08-31), 232028(2026-10-19), 240129(2026-11-30), 240278(2026-11-30), 240557(2027-01-22), 240746(2027-01-22), 240794(2027-02-28), 241077(2027-04-08), 241188(2027-04-08), 241462(2027-05-03), 241523(2027-05-03), 241767(2027-06-26), 241910(2027-06-26), 242159(2027-08-09), 242322(2027-09-10), 242518(2027-09-10), 242694(2027-10-25), 242695(2027-10-25)
510(k) numbers
["K172315","K211738","K242757"]
PMA numbers
not on file
Product code
MCW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register