FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: 180-Gram Cement Cartridge with Breakaway Femoral Nozzle, Product Number: 0206-530-000, UDI: (01)4546540857880 for orthopedic use.
- Recall number
- Z-1244-2018
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- 180-Gram Cement Cartridge with Breakaway Femoral Nozzle, Product Number: 0206-530-000, UDI: (01)4546540857880 for orthopedic use.
- Status
- Terminated
- Initiated
- 2017-11-09
- Posted
- not on file
- Terminated
- 2019-04-16
- Reason
- Lack of sterility assurance as bioburden levels were found to be higher than internal acceptable rates.
- Root cause
- Environmental control
- Action
- Letters dated November 16, 2017, were sent to customers. Actions to be taken by the Customer/User: 1. Immediately review this Recall Notification. 2. Check all stock areas and/or operating room storage to determine if any devices from the affected lots are at your facility. 3. Quarantine and discontinue use of the recalled devices. 4. Complete and sign the enclosed Business Reply Form (BRF). 5. If you have further distributed these products, please forward this letter and the attached BRF to all affected locations. Please indicate each location on the BRF. 6. Return the completed BRF to Stryker Instruments Post Market Regulatory Compliance via fax (866-521- 2762) or email (erin.bissonnette@stryker.com). 7. If the BRF for your facility indicates that recalled product is currently on hand, a FedEx label will be emailed to you. This shipping label should be used to return recalled product. Upon receipt of the recalled product, a credit will be issued to your account. 8. Report any adverse events associated with the use of these devices to Stryker Instruments 1-800-253-3210, including any infections potentially associated with the use of affected product.
- Quantity
- 708
- Distribution
- Distributed in all 50 U.S. States and the District of Columbia. Distributed in Australia, Argentina, Japan, Canada, Chile, Korea, Spain, France, Italy, Netherlands, and United Kingdom.
- Lot, serial or model codes
- Lot Numbers: 17277012, 17283012 UDI: (01)4546540857880
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JDZ
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28