Skip to the content

FDA Medical Device Recalls (openFDA)

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK: MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.

Recall number
Z-1244-2016
Recalling firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Product
MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.
Status
Terminated
Initiated
2015-12-15
Posted
2016-03-24
Terminated
2017-04-20
Reason
Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" under specific conditions.
Root cause
Software design
Action
Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" if all the following conditions are met: (1) At least one patient has been treated on the day; (2) A major error occurs when VeroTM/MHI-TM2000 is operated in a mode, other than Clinical Mode, which means Daily Check Mode or Maintenance Mode. (3) The major error is NOT associated with a function that uses therapeutic/ MV radiation parameters, such as an error in Isocenter Check function, etc. Note 1: If the major error is associated with a function that uses therapeutic/MV radiation parameters, the problem that treatment completion status is changed will never occur. The Operator Console changes the treatment completion status in the course of processing the major error. Note 2: Although treatment completion status is changed to "Untreated", its delivered dose result has been correctly saved into R&V System. It has not been lost. Advice on Actions by Customer/Users: 1. Please make sure to perform the operations as described below in case a major error occurred in a mode other than Clinical Mode. (1) Cancel the error. (2) Click [Clinical] button on the left of Operator Console screen. (3) Verify, on R&V System screen, that treatment completion status of the patients who were treated on the day are all correct. The treatment completion status is indicated in different colors as follows depending on the status: Untreated: BLUE, Completed: GRAY, Discontinued: RED. Customers requiring support concerning this issue, please contact the Customer Service Representative: 1-800-597-5911 (Mon - Fri) or via email us.support@brainlab.com. If any other problem is detected, please stop operation and contact the Customer Service representative.
Quantity
25 devices (4 domestically in U.S.A. and 21 internationally)
Distribution
Distributed in the states of Florida, New York, Ohio & Texas, and the countries of France, Germany, Japan, Italy, Korea, & Belgium.
Lot, serial or model codes
Serial numbers: 201901, 201902, 201903, 201904, 201905, 202901, 202902, 202903, 202904, 202905, 202906, 203901, 203902, 203903, 203904, 203905, 203906, 203918, 203919, 203921, 203922, 203924, 203925, 203927 and 203929.
510(k) numbers
["K072047","K122450","K152867"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3006942329
Firm city
HIROSHIMA
Firm state
4-CHOME KAN-ON-SHIN-MACHI
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register