Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Part 475630, Intramedullary Bone Saw, Cam Assembly Size 14 mm, The Biomet Intramedullary Bone Saw is a unique and exclusive clinical instrument that is used to correct leg length, or rotational discrepancies in the femur.

Recall number
Z-1244-2014
Recalling firm
Biomet, Inc.
Product
Part 475630, Intramedullary Bone Saw, Cam Assembly Size 14 mm, The Biomet Intramedullary Bone Saw is a unique and exclusive clinical instrument that is used to correct leg length, or rotational discrepancies in the femur.
Status
Terminated
Initiated
2014-02-26
Posted
2014-03-20
Terminated
2015-07-07
Reason
Incorrect labeling on specific lots of the Intramedullary Bone Saw. The Saw Blade Assembly was identified as CAM Assembly.
Root cause
Labeling False and Misleading
Action
The firm, BIOMET, sent an "URGENT MEDICAL DEVICE SAFETY ALERT" letter dated February 26, 2014 via Fed Ex overnight to domestic consignees. Australia and GSCC were notified by email and also received a hard copy letter via mail. The letter described the product, problem and actions to be taken. The letter stated that this action requires the immediate location and discontinued use of the product and its return to Biomet. The Consignees were were instructed to immediately locate, discontinue use of the product and remove the identified device(s) listed; carefully follow the instructions on the FAX Back Response Form and reply within three (3) business days; complete and return Form via Fax to: 574-372-1683 prior to return of product, and use priority carrier for your shipment. If you have further distributed this product, you MUST notify hospital personnel of this action via enclosed "Dear Risk/Recall Manager Notice", and note that you are charged with the location and return of these products. Shipping address is Return Goods, 56 E. Bell Drive Warsaw, IN 46582. Questions related to the notice should be directed to 574-372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity
6
Distribution
Worldwide Distribution: US (nationwide) to states of: NY, TX, MI, and M; to countries of: Australia and The Netherlands.
Lot, serial or model codes
Lot 239470, 457450
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register