FDA Medical Device Recalls (openFDA)
Migada Inc.: TEVADAPTOR Closed Drug Reconstitution and Transfer System, a closed system used to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection. Model number is MG412110. The device is a single use device.
- Recall number
- Z-1244-2009
- Recalling firm
- Migada Inc.
- Product
- TEVADAPTOR Closed Drug Reconstitution and Transfer System, a closed system used to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection. Model number is MG412110. The device is a single use device.
- Status
- Terminated
- Initiated
- 2008-12-18
- Posted
- 2009-05-06
- Terminated
- 2009-07-27
- Reason
- Syringe adaptor is disintegrating when attached to the end of the syringe.
- Root cause
- Component design/selection
- Action
- A Field Safety Notice, dated 12/8/2008, was sent to all TEVADAPTOR customers advising them not to use drugs containing N,N-Dimethylacetamide. Teva Medical is presently revising the instruction leaflet supplied with each TEVADAPTOR to include a warning against the use of drugs containing N,N-Dimethylacetamide. Teva Medical will introduce this revised leaflet to current production as soon as practicable. In the meantime, Teva Medical will advise all users of the potential problem by means of the Urgent Field Safety Notice. Customers are to ensure that the safety notice is passed to all those who need to be aware within their organizations. Questions should be directed to their local contacts at B Braun, USA.
- Quantity
- 900,000 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- all lots
- 510(k) numbers
- ["K051669","K071741"]
- PMA numbers
- not on file
- Product code
- LHI
- Firm FEI
- 3002807615
- Firm city
- Kiryat Shmona
- Firm state
- not on file
- Firm country
- Israel
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28