Abbott Laboratories MPG: CA 15-3 Controls, list 9C08-10; an In Vitro diagnostic for use with the AxSYM or IMx CA 15-3 reagents, calibrators and instruments; each kit contains 2 bottles with 115D8:DF3 reactive determinants (human) prepared in TRIS buffer with protein (bovine) stabilizers to yield the following assay values (U/ml): Control L - 35 and Control H - 150; Abbott Laboratories, Abbott Park, IL 60064
- Recall number
- Z-1244-06
- Recalling firm
- Abbott Laboratories MPG
- Product
- CA 15-3 Controls, list 9C08-10; an In Vitro diagnostic for use with the AxSYM or IMx CA 15-3 reagents, calibrators and instruments; each kit contains 2 bottles with 115D8:DF3 reactive determinants (human) prepared in TRIS buffer with protein (bovine) stabilizers to yield the following assay values (U/ml): Control L - 35 and Control H - 150; Abbott Laboratories, Abbott Park, IL 60064
- Status
- Terminated
- Initiated
- 2006-06-09
- Posted
- 2006-07-18
- Terminated
- 2008-01-18
- Reason
- The CA 15-3 Controls replicates out of range high resulting in invalid runs.
- Root cause
- Other
- Action
- Abbott initiated a field correction recall by letter dated 6/9/06 and sent on 6/12/06 to all customers receiving the affected controls. The accounts were informed of the increase in invalid runs due to controls outside of the package insert range with this lot of CA 15-3 Controls. The accounts were requested to assess the inventory for the affected lot of controls, discontinue use of the lot and destroy the remaining inventory of the lot, recording the number of kits on the customer reply form and faxing it to Abbott by 6/21/06, indicating that the letter was received and understood. Credit will be issued for the number of kits destroyed, as indicated on the reply sheet. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.
- Quantity
- 1794 kits
- Distribution
- Worldwide, including U.S.A., Puerto Rico, Canada, Mexico, Ecuador, Chile, Argentina, Brazil, Germany, Japan, SIngapore, Hong Kong, Thailand, South Korea and Australia.
- Lot, serial or model codes
- list 9C08-10; lot number 29506M100, expiration date July 5, 2006
- 510(k) numbers
- ["K963926","K964407"]
- PMA numbers
- not on file
- Product code
- MOI
- Firm FEI
- 1415939
- Firm city
- Abbott Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28