Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc: IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders

Recall number
Z-1243-2016
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders
Status
Terminated
Initiated
2016-01-13
Posted
not on file
Terminated
2018-01-08
Reason
The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
Root cause
Device Design
Action
Siemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Customers where instructed to complete and return the Correction Effectiveness Check attached. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity
9334 units
Distribution
Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
Lot, serial or model codes
All lots
510(k) numbers
["K932926"]
PMA numbers
not on file
Product code
CHP
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register