FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc. dba Mindray North America: Gas Module 3 monitoring spirometer, Part Number: 115-030108-00 Product Usage: The Gas Module 3 is an accessory device designed to monitor airway gases during anesthesia and/or assisted respiration. Monitored gases includes O2, CO2, N2O and anesthetic agents Halothane, Isoflurane, Sevoflurane, Desflurane and Enflurane. It is intended for use in a fixed location such as the operating room or intensive care unit. The Gas Module 3 is compatible with Mindray patient monitors which display the gas analysis results through the use of a proprietary communication protocol.

Recall number
Z-1242-2018
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Product
Gas Module 3 monitoring spirometer, Part Number: 115-030108-00 Product Usage: The Gas Module 3 is an accessory device designed to monitor airway gases during anesthesia and/or assisted respiration. Monitored gases includes O2, CO2, N2O and anesthetic agents Halothane, Isoflurane, Sevoflurane, Desflurane and Enflurane. It is intended for use in a fixed location such as the operating room or intensive care unit. The Gas Module 3 is compatible with Mindray patient monitors which display the gas analysis results through the use of a proprietary communication protocol.
Status
Terminated
Initiated
2017-12-01
Posted
not on file
Terminated
2018-08-29
Reason
The firm made multiple changes to the product requiring the submission of a new 510(k) premarket notification.
Root cause
Device Design
Action
Mindray sent an Safety Notification letter dated December 1, 2017 to all customers. The letter identified the affected product, problem and actions to be taken. For question call 201-995-8407.
Quantity
1719
Distribution
Worldwide Distribution - U.S. Nationwide in the states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY, including Puerto Rico and distributed to 30 foreign consignees.
Lot, serial or model codes
All Gas Modules 3 with Part Number 115-030108-00 purchased between February 14, 2008 and October 2015.
510(k) numbers
["K062754"]
PMA numbers
not on file
Product code
BZK
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register