Skip to the content

FDA Medical Device Recalls (openFDA)

Richard Wolf Medical Instruments Corp.: The I-Cart is used to store, transport, and power video equipment used in medical procedures.

Recall number
Z-1240-2015
Recalling firm
Richard Wolf Medical Instruments Corp.
Product
The I-Cart is used to store, transport, and power video equipment used in medical procedures.
Status
Terminated
Initiated
2014-12-10
Posted
2015-03-03
Terminated
2017-07-31
Reason
The transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.
Root cause
Device Design
Action
Field Action Notice letters dated December 10, 2014 were sent to both direct accounts (customers). The letters instructed customers to: 1) Instruct the biomed to bypass the integrated power grid on the cart and not to use the electrical system on the affected carts until a qualified Omni Corporation service technician is able to inspect and repair the carts; and 2) Omni Corporation will schedule a qualified Omni field service technician, in conjunction with Richard Wolf and the customer's biomed unit, to visit and replace the transformer on the cart.
Quantity
83 carts
Distribution
Nationwide
Lot, serial or model codes
(NOTE: The listed serial numbers are for the carts manufactured and include those that were distributed.)   1) Part Number: 31113.801; Serial Numbers: 1306000, 1306007, 1306014, 1306021, 1306103, 1306001, 1306008, 1306015, 1306022, 1306104, 1306002, 1306009, 1306016, 1306023, 1306105, 1306003, 1306010, 1306017, 1306024, 1307100, 1306004, 1306011, 1306018, 1306100,  1307101, 1306005, 1306012, 1306019, 1306101, 1307102, 1306006, 1306013, 1306020, 1306102, 1317103.   2) Part Number: 31113.701; Serial Numbers: 11001, 11011, 11020, 11030, 11039, 11048, 11058, 11002, 11012, 11021, 11031, 11040, 11049, 11059, 11003, 11013, 11023, 11032, 11041, 11050, 11060, 11005, 11014, 11024, 11033, 11042, 11051, 11061, 11006, 11015, 11025, 11034, 11043, 11052, 11063, 11007, 11016, 11026, 11035, 11044, 11053, 11064, 11008, 11017, 11027, 11036, 11045, 11055, 11065, 11009, 11018, 11028, 11037, 11046, 11056, 11066, 11010, 11019, 11029, 11038, 11047, 11057, 11067.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KMI
Firm FEI
1418479
Firm city
Vernon Hills
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register