FDA Medical Device Recalls (openFDA)
Convatec Inc.: 'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL Product Usage: Latex Lubricous Coated Foley Catheters The standard catheter essentially consists of a flexible lubricous coated latex tube which incorporates one major drainage lumen and one or two minor lumens constructed in the wall of the tube. One minor lumen is for inflating the retention balloon (available in 2-Way & 3-Way catheters). The second minor lumen (applicable for 3 Way catheters only) is used for bladder irrigation purposes.
- Recall number
- Z-1240-2014
- Recalling firm
- Convatec Inc.
- Product
- 'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL Product Usage: Latex Lubricous Coated Foley Catheters The standard catheter essentially consists of a flexible lubricous coated latex tube which incorporates one major drainage lumen and one or two minor lumens constructed in the wall of the tube. One minor lumen is for inflating the retention balloon (available in 2-Way & 3-Way catheters). The second minor lumen (applicable for 3 Way catheters only) is used for bladder irrigation purposes.
- Status
- Terminated
- Initiated
- 2014-02-07
- Posted
- 2014-03-20
- Terminated
- 2014-12-18
- Reason
- The lumen of the Medline Sterile Foley Catheter was found to be blocked by excess lubricious coating material.
- Root cause
- Device Design
- Action
- Convatec sent Urgent: Voluntary Medical Device Recall letters dated 2/7/2014 via UPS and email. Foreign customers were provided the notice via DHL. The letter identified the affected product, problem and actions to be taken. Customers are asked to perform a count of recalled product in inventory. All consignees should be contacted and be instructed to return product based on their procedures. Customers were instructed to complete and return the enclosed respoonse form.
- Quantity
- 54,590 catheters
- Distribution
- Worldwide Distribution and in the state of TX
- Lot, serial or model codes
- 540204R001
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KOD
- Firm FEI
- 3011987967
- Firm city
- Skillman
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as