Smiths Medical PM, Inc.: BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
- Recall number
- Z-1240-2007
- Recalling firm
- Smiths Medical PM, Inc.
- Product
- BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
- Status
- Terminated
- Initiated
- 2007-07-18
- Posted
- 2007-09-21
- Terminated
- 2008-10-31
- Reason
- Phantom pacer spikes caused by a damaged resistor.
- Root cause
- Other
- Action
- Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.
- Quantity
- 178
- Distribution
- World wide: USA including states of KY, MS, PA, LA, and PR and countries of Poland, Mexico, Kuwait, Taiwan, Singapore, Hon Kong, India, Bolivia, Colombia, Chile, Argentia, Brunei Darssalam, Israel, Greece, Romania, Germany, The Netherlands, France, Great Britain, Spain, Brazil, Russia, China.
- Lot, serial or model codes
- Catalog number/ serial number: 3404001/AI05110052/ AI05120000/ AI05120011/ AI05120012/ AI05120013/ AI05120018/ AI06060015/ AI06070047/ AI06100004/ AI06100005/ AI06100006/ , 3404001S/AI05110000/ AI05110001/ AI05110002/ AI05110003/ AI05110004/ AI05110005/ AI05110006/ AI05110007/ AI05110008/ AI05110009/ AI05110010/ AI05110011/ AI05110012/ AI05110013/ AI05110014/ AI05110015/ AI05110016/ AI05110017/ AI05110018/ AI05110019/ AI05110020/ AI05110021/ AI05110022/ AI05110023/ AI05110024/ AI05110025/ AI05110026/ AI05110027/ AI05110028/ AI05110029/ AI05110030/ AI05110031/ AI05110032/ AI05110033/ AI05110034/ AI05110035/ AI05110036/ AI05110037/ AI05110038/ AI05110039/ AI05120010/ AI05120016/ AI05120017/ AI05120019/ AI06010042/ AI06010043/ AI06060003/ AI06060004/ AI06060005/ AI06060006/ AI06060007/ AI06060008/ AI06060009/ AI06060010/ AI06060011/ , 3404002/AI05090021/ AI05090024/ AI05110049/ AI05120015/ AI06010035/ AI06010036/ AI06010049/ AI06040000/ AI06040001/ AI06040002/ AI06040003/ AI06040004/ AI06040005/ AI06040006/ AI06040007/ AI06040008/ AI06040009/ AI06040010/ AI06040011/ AI06040012/ AI06040013/ AI06040014/ AI06040015/ AI06040016/ AI06040017/ AI06040018/ AI06040020/ AI06060012/ AI06060013/ AI06060014/ AI06060017/ AI06060018/ AI06060019/ AI06060020/ AI06070037/ AI06070038/ AI06070039/ AI06070040/ AI06070042/ AI06070043/ AI06070044/ AI06070045/ AI06090053/ AI06100011/ AI06100012/ AI07010057/ AI07010058/ , 3404002E/AI05110040 AI05110041/ AI05110042/ AI05120001/ AI05120002/ AI05120005/ AI05120006/ AI05120007/ AI05120008/ AI05120009/ AI05120024/ AI06010041/ AI06010044/ AI06010045/ AI06010046/ AI06010048/ AI06040019/ AI06040023/ AI06040024/ AI06070041/ AI06070046/ AI06070048/ AI06070049/ AI06090050/ AI06090051/ , 3404002F/AI07010067/ AI07010068/ , 3404002G/AI05110050/ AI05110051/ AI05120003/ AI05120004/ AI06060021/ AI06060022/ AI06060023/ AI06070035/ AI06070036/ AI06090054/ AI06090055/ AI06090056/ AI06090057/ AI06090058/ AI06100008/ AI06100009/ AI06100010/ AI06100018/ , 3404002S/AI05110043/ AI05110044/ AI05110045/ AI06010037/ AI06010038/ AI06010039/ AI06010040/ AI06060000/ AI06060001/ AI06060002/ AI06060016/ AI06090052/ AI06100019/ AI06100020/ AI06100021/ AI06100022/ AI06100023/ AI06100024/ AI07010050/ AI07010051/
- 510(k) numbers
- ["K981939"]
- PMA numbers
- not on file
- Product code
- DQA
- Firm FEI
- 2183502
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28