Smiths Medical PM, Inc.: BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc

Recall number
Z-1240-2007
Recalling firm
Smiths Medical PM, Inc.
Product
BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
Status
Terminated
Initiated
2007-07-18
Posted
2007-09-21
Terminated
2008-10-31
Reason
Phantom pacer spikes caused by a damaged resistor.
Root cause
Other
Action
Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.
Quantity
178
Distribution
World wide: USA including states of KY, MS, PA, LA, and PR and countries of Poland, Mexico, Kuwait, Taiwan, Singapore, Hon Kong, India, Bolivia, Colombia, Chile, Argentia, Brunei Darssalam, Israel, Greece, Romania, Germany, The Netherlands, France, Great Britain, Spain, Brazil, Russia, China.
Lot, serial or model codes
Catalog number/ serial number: 3404001/AI05110052/ AI05120000/ AI05120011/ AI05120012/ AI05120013/ AI05120018/ AI06060015/ AI06070047/ AI06100004/ AI06100005/ AI06100006/ , 3404001S/AI05110000/ AI05110001/ AI05110002/ AI05110003/ AI05110004/ AI05110005/ AI05110006/ AI05110007/ AI05110008/ AI05110009/ AI05110010/ AI05110011/ AI05110012/ AI05110013/ AI05110014/ AI05110015/ AI05110016/ AI05110017/ AI05110018/ AI05110019/ AI05110020/ AI05110021/ AI05110022/ AI05110023/ AI05110024/ AI05110025/ AI05110026/ AI05110027/ AI05110028/ AI05110029/ AI05110030/ AI05110031/ AI05110032/ AI05110033/ AI05110034/ AI05110035/ AI05110036/ AI05110037/ AI05110038/ AI05110039/ AI05120010/ AI05120016/ AI05120017/ AI05120019/ AI06010042/ AI06010043/ AI06060003/ AI06060004/ AI06060005/ AI06060006/ AI06060007/ AI06060008/ AI06060009/ AI06060010/ AI06060011/ , 3404002/AI05090021/ AI05090024/ AI05110049/ AI05120015/ AI06010035/ AI06010036/ AI06010049/ AI06040000/ AI06040001/ AI06040002/ AI06040003/ AI06040004/ AI06040005/ AI06040006/ AI06040007/ AI06040008/ AI06040009/ AI06040010/ AI06040011/ AI06040012/ AI06040013/ AI06040014/ AI06040015/ AI06040016/ AI06040017/ AI06040018/ AI06040020/ AI06060012/ AI06060013/ AI06060014/ AI06060017/ AI06060018/ AI06060019/ AI06060020/ AI06070037/ AI06070038/ AI06070039/ AI06070040/ AI06070042/ AI06070043/ AI06070044/ AI06070045/ AI06090053/ AI06100011/ AI06100012/ AI07010057/ AI07010058/ , 3404002E/AI05110040 AI05110041/ AI05110042/ AI05120001/ AI05120002/ AI05120005/ AI05120006/ AI05120007/ AI05120008/ AI05120009/ AI05120024/ AI06010041/ AI06010044/ AI06010045/ AI06010046/ AI06010048/ AI06040019/ AI06040023/ AI06040024/ AI06070041/ AI06070046/ AI06070048/ AI06070049/ AI06090050/ AI06090051/ , 3404002F/AI07010067/ AI07010068/ , 3404002G/AI05110050/ AI05110051/ AI05120003/ AI05120004/ AI06060021/ AI06060022/ AI06060023/ AI06070035/ AI06070036/ AI06090054/ AI06090055/ AI06090056/ AI06090057/ AI06090058/ AI06100008/ AI06100009/ AI06100010/ AI06100018/ , 3404002S/AI05110043/ AI05110044/ AI05110045/ AI06010037/ AI06010038/ AI06010039/ AI06010040/ AI06060000/ AI06060001/ AI06060002/ AI06060016/ AI06090052/ AI06100019/ AI06100020/ AI06100021/ AI06100022/ AI06100023/ AI06100024/ AI07010050/ AI07010051/
510(k) numbers
["K981939"]
PMA numbers
not on file
Product code
DQA
Firm FEI
2183502
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register