Skip to the content

FDA Medical Device Recalls (openFDA)

Aksys, Ltd.: Aksys PHD Personal Hemodialysis System for daily home hemodialysis; an automated high permeability hemodialysis system, water purification system for hemodialysis; Manufactured in the USA for Aksys Ltd., Lincolnshire, IL 60069 USA; Model 1M101

Recall number
Z-1239-06
Recalling firm
Aksys, Ltd.
Product
Aksys PHD Personal Hemodialysis System for daily home hemodialysis; an automated high permeability hemodialysis system, water purification system for hemodialysis; Manufactured in the USA for Aksys Ltd., Lincolnshire, IL 60069 USA; Model 1M101
Status
Terminated
Initiated
2006-06-26
Posted
2006-07-13
Terminated
2008-05-09
Reason
The hemodialysis device was being marketed with a modified treatment length that exceeds the treatment length for which the device was originally designed.
Root cause
Other
Action
Aksys sent a Clinical Bulletin: Field Correction Recall, CB Number: 048, dated 6/26/2006, via overnight Federal Express to all clinics or patients currently using the PHD instrument. CB #048 advises patients that until field modifications to the PHD System can be performed to limit the maximum treatment time to 170 minutes, Aksys recommends that users not exceed this treatment time in the interim. Each user was requested to contact his/her clinic to discuss how this change may affect treatment options and to complete and mail an enclosed pre-paid, self-addressed receipt acknowledgement card. A revised PHD System Operatror''s Manual was issued to all PHD System users on 7/12/06, which reflects the reduction in the maximum treatment time.
Quantity
133 units
Distribution
Nationwide (California, Connecticut, Florida, Georgia, Illinois, Maryland, Massachusetts, Michigan, Ohio, Nebraska, New Jersey, New York, Tennessee, Texas, Virginia, Washington and Wisconsin), and internationally to the United Kingdom.
Lot, serial or model codes
Serial numbers A000089, A000093, A000095, A100002, A100003, A100005, A100009, A100016, A100024, A100031, A100042, A100043, A100049, A100066, A100067, A100076, A100077, A100081, A100083, A100084, A100094, C100104, C100107, C100109, C100114, C100116, C100117, C100123, C100124, C100125, C100126, C100127, C100128, C100131, C100132, C100133, C100136, C100139, C100140, C100142, C100143, C100146, C100149, C100151, C100152, C100157, C100161, C100162, C100163, C100164, C100166, C100173, C100174, C100175, C100176, C100178, C100179, C100180, C100181, C100185, C100187, C100188, C100190, C100191, C100192, C100194, C100195, C100198, C100199, C100200, C100201, C100202, C100204, C100206, C100207, C100209, C100211, C100212, C100214, C100217, C100218, C100219, C100222, C100226, C100227, C100228, C100229, C100230, C100232, C100233, C100234, C100236, C100238, C100239, C100240, C100241, C100243, C100244, C100245, C100246, C100247, C100248, C100249, C100251, C100252, C100253, C100254, C100255, C100256, C100257, C100259, C100260, C100263, C100264, C100265, C100266, C100302, C100305, C100306, C100307, C100308, C100310, C100311, C100312, C100314, C100315, C100317, C100318, C100320, C100321, C100326, C100328, C100329.
510(k) numbers
["K010131"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000265533
Firm city
Lincolnshire
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register