Baxa Corporation: Exacta-Mix 2400, OS v1.07, Model No. 8300-0073, Pharmacy Compounding System, Baxa Corporation, Englewood, CO 80112.

Recall number
Z-1238-2007
Recalling firm
Baxa Corporation
Product
Exacta-Mix 2400, OS v1.07, Model No. 8300-0073, Pharmacy Compounding System, Baxa Corporation, Englewood, CO 80112.
Status
Terminated
Initiated
2007-06-27
Posted
2007-09-27
Terminated
2007-12-07
Reason
A software defect could allow up to 50mL of extra volume being added to a TPN solution.
Root cause
Other
Action
All customers were notified via telephone on 06/27/2007 with follow-up of a faxed/emailed Urgent Product Recall letter dated 27 June 2007, detailing the software defect. End users were given a work around procedure to use until conversion back to version 1.06 could be scheduled. An additional copy of the recall letter was sent by FedEx.
Quantity
30 units
Distribution
Class I Recall-Worldwide Distribution: USA including states of CA, CO, FL, KY, MI, NY, OH, OR, and TX and the UK
Lot, serial or model codes
Version 1.07 of the EM2400 Operating Software. (Serial Numbers: 36805,36806,36356,36399,36804,36799,36797,36398,32438,36798,36460,36794,32346,36800,36801,36802,36803,(37767-36773), (36788-36793),36795 and 36796 (foreign)
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1419106
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register