Tornier, Inc: TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

Recall number
Z-1237-2026
Recalling firm
Tornier, Inc
Product
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
Status
Open, Classified
Initiated
2025-12-19
Posted
2026-02-02
Terminated
not on file
Reason
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
Root cause
Labeling mix-ups
Action
On December 19, 2025, the firm began notifying affected customers via Urgent Field Safety Notice letters. Customers were instructed to immediately check their internal inventory for affected product. Firm stated that return of product is not necessary because the issue is detectable pre-operatively. The physical product itself is correctly marked, and after unpacking, should only be assigned to the correct place in the screw caddy. The notification is being provided for awareness only.
Quantity
427
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot AZ4824290
510(k) numbers
["K161742"]
PMA numbers
not on file
Product code
KWS
Firm FEI
3000718467
Firm city
Bloomington
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register