Skip to the content

FDA Medical Device Recalls (openFDA)

PerkinElmer Life and Analytical Sciences, Wallac, OY: DELFIA hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.

Recall number
Z-1237-2019
Recalling firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Product
DELFIA hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.
Status
Open, Classified
Initiated
2019-01-10
Posted
not on file
Terminated
not on file
Reason
hCG analyte concentrations may be decreased up to 15- 20% at the lower measuring range of hCG concentrations.
Root cause
Nonconforming Material/Component
Action
Perkin Elmer initiated notification on January 10, 2018. The Recall Letter delivered to the end customer by email facilitated through the PerkinElmer Technical Support Group. PerkinElmer requires customers to: Inspect DELFIA / AutoDELFIA hCG kit inventory for the affected lots. Review your QC follow-up of the AutoDELFIA hCG assay for potential decrease in the control concentrations. If the control values are within your established acceptance range it is safe to report the results To avoid unnecessary repeat testing of samples, if the control values are out of the acceptance range discontinue use of the DELFIA / AutoDELFIA hCG kit lot and dispose the remaining kits of the lot. Complete the Response Form with the quantity of affected DELFIA / AutoDELFIA hCG kits you have disposed from your inventory
Quantity
31 kits (10 in US, 21 in ROW)
Distribution
Worldwide distribution - US Nationwide distribution. and countries of Austria, Bolivia, UK, Switzerland,Thailand, Finland.
Lot, serial or model codes
Kit Lot Numbers: 664891, 666471, 666604 Pack Lot Numbers: 1066489101, 1066647101,1066660401 UDI:  (01)06438147009176(17)190430(10)664891 (01)06438147009176(17)190430(10)666471 (01)06438147009176(17)190430(10)666604
510(k) numbers
["K862044"]
PMA numbers
not on file
Product code
JHJ
Firm FEI
3002808498
Firm city
Turku
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register