Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Solar HA Humeral Stem with Plasma Coating 7mm to 17 mm Product Catalog Number: 5351-4307 to 5351-4317. For use as a Shoulder Replacement. Stryker Orthopaedics Made in USA

Recall number
Z-1236-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Solar HA Humeral Stem with Plasma Coating 7mm to 17 mm Product Catalog Number: 5351-4307 to 5351-4317. For use as a Shoulder Replacement. Stryker Orthopaedics Made in USA
Status
Terminated
Initiated
2007-11-15
Posted
2008-06-26
Terminated
2010-01-25
Reason
Delamination: Loosening of the titanium plasma coating on the product was observed on twelve non-distributed units.
Root cause
Process control
Action
On 11/28/2007, Stryker Orthopaedics sent, via FedEx, Urgent Product Recall notification letters to distributors, hospitals, surgeons and hospitals' Chiefs of Orthopaedics informing them of the firm's voluntary recall of the affected product. They were instructed to discontinue use of the product and to have the units returned to the firm. Distributors were further instructed to notify their customers using the sub-recall letter provided by the firm. The letter to surgeons also informed them to monitor their patients who had received an implant of the product.
Quantity
915 units
Distribution
Nationwide - USA
Lot, serial or model codes
All Lot Codes for the following: Catalog # 5351-4307, 7mm; 5351-4308, 8mm; 5351-4309, 9mm; 5351-4310, 10mm; 5351-4311, 11mm; 5351-4312, 12mm; 5351-4313, 13mm; 5351-4314, 14mm; 5351-4315, 15mm; 5351-4316, 16mm; and 5351-4317, 17mm.
510(k) numbers
["K061677"]
PMA numbers
not on file
Product code
KWS
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register