FDA Medical Device Recalls (openFDA)
Sensus Healthcare, Inc.: Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities
- Recall number
- Z-1235-2023
- Recalling firm
- Sensus Healthcare, Inc.
- Product
- Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities
- Status
- Terminated
- Initiated
- 2023-02-10
- Posted
- 2023-03-08
- Terminated
- 2024-01-31
- Reason
- When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.
- Root cause
- Software design
- Action
- Sensus healthcare issued an "Urgent Device Correction" to its consignees on 02/10/2023 by email. The notice requested the following actions: "Do not initiate beam delivery after a pause & resume sequence without verifying the amount of dose delivered. Do evaluate any patient treatments with version 1.9.2 that followed the pause & resume workflow. Do verify that the amount of dose to be delivered is as expected after each pause & resume sequence." The firm has developed a permanent software resolution for this issue. Customer service will contact the customer to coordinate upgrading to this revised software. Questions: Contact Sensus Customer Support at support@sensushealthcare.com or 561-578-4982
- Quantity
- 23 units
- Distribution
- US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
- Lot, serial or model codes
- Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272
- 510(k) numbers
- ["K150037"]
- PMA numbers
- not on file
- Product code
- JAD
- Firm FEI
- 3008513398
- Firm city
- Boca Raton
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28