FDA Medical Device Recalls (openFDA)

Smiths Medical ASD, Inc.: CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112.

Recall number
Z-1235-2010
Recalling firm
Smiths Medical ASD, Inc.
Product
CADD-LEGACY 1 Model 6400, Ambulatory Infusion Pump, REF 21-6400-51, Rx Only, Smiths Medical MD, Inc.., St. Paul, MN 55112.
Status
Terminated
Initiated
2009-12-23
Posted
2010-04-05
Terminated
2011-12-10
Reason
The CADD-Legacy 1 Model 6400 Ambulatory Infusion Pump (Serial Number 370969) was shipped without documentation that the required accuracy testing had been performed during the manufacturing process. If the subject infusion pump did not go through the accuracy testing, and if it does not meet accuracy specifications, there is a potential for over or under-delivery.
Root cause
Employee error
Action
Smiths Medical telephoned the customer on December 23, 2009, to return the CADD-Legacy 1 Ambulatory Infusion Pump, serial number 370969, since there was no record of the completed accuracy testing for the infusion pump.
Quantity
1
Distribution
MI
Lot, serial or model codes
Serial number: 370969
510(k) numbers
["K982838"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register