Siemens Medical Solutions USA, Inc: syngo Imaging, picture archiving and communications system (PACS), Model number 10 014 063
- Recall number
- Z-1235-2009
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- syngo Imaging, picture archiving and communications system (PACS), Model number 10 014 063
- Status
- Terminated
- Initiated
- 2009-01-12
- Posted
- 2009-04-28
- Terminated
- 2009-09-23
- Reason
- Incorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image availability, inaccurate results during image measurements, incorrect review or diagnosis of displayed or printed images, measurement inaccuracies from the host modality and image
- Root cause
- Software design
- Action
- syngo Imaging has issued an Addendum to the User Manual, which was distributed to affected customers via Update Instructions IM118/08/S in November, 2008. Two safety related hotfixes (VB20G_HF4 and VB30A_HF26) will be released to those affected only after Update Instructions IM118/08/S has been performed. The addendum informs the user about safety relevant notes which pertain to syngo Imaging versions VB20G and VB30A. Update is a prerequisite for the following safety related hotfixes: VB20G_HF4 and VB30A_HF26.
- Quantity
- 55 units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Model number 10 014 063. Serial numbers: 1005, 1039, 1040, 1041, 1043, 1049, 1053, 1055, 1056, 1057, 1058, 1061, 1064, 1065, 1067, 1071, 1074, 1076, 1079, 1080, 1081, 1082, 1083, 1084, 1086, 1090, 1105, 1107, 1112, 1114, 1116, 1119, 1124, 1174, 1188, 1189, 1202, 1203, 1210, 1218, 1226, 1227, 1229, 1240, 1246, 1247, 1258, 1269, 1271, 1278, 1279, 1310, 1316, 1329, and 1334.
- 510(k) numbers
- ["K071114"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28