Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: ACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM STERILE ATN LAG SCREW 105MM STERILE ATN LAG SCREW 110MM STERILE ATN LAG SCREW 115MM STERILE ATN LAG SCREW 120MM STERILE ATN LAG SCREW 125MM STERILE ATN LAG SCREW 130MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Recall number
Z-1234-2018
Recalling firm
Zimmer Biomet, Inc.
Product
ACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM STERILE ATN LAG SCREW 105MM STERILE ATN LAG SCREW 110MM STERILE ATN LAG SCREW 115MM STERILE ATN LAG SCREW 120MM STERILE ATN LAG SCREW 125MM STERILE ATN LAG SCREW 130MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
Status
Terminated
Initiated
2017-12-20
Posted
2018-03-14
Terminated
2020-04-22
Reason
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
Root cause
Device Design
Action
On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.
Quantity
3454
Distribution
USA (nationwide)
Lot, serial or model codes
903311060 903311065 903311070 903311075 903311080 903311085 903311090 903311095 903311100 903311105 903311110 903311115 903311120 903311125 903311130
510(k) numbers
["K010780","K013563","K042325","K111663"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register