Skip to the content

FDA Medical Device Recalls (openFDA)

Zipline Medical: Zip 8i Surgical Skin Closure Device; for incisions up to 8 cm. Product number PS2080 Indicated for use during and after skin incision procedures to approximate skin and hold together skin edges until healing can take place.

Recall number
Z-1234-2015
Recalling firm
Zipline Medical
Product
Zip 8i Surgical Skin Closure Device; for incisions up to 8 cm. Product number PS2080 Indicated for use during and after skin incision procedures to approximate skin and hold together skin edges until healing can take place.
Status
Terminated
Initiated
2015-02-05
Posted
2015-03-02
Terminated
2015-04-10
Reason
The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility.
Root cause
Packaging process control
Action
Zipline sent an Urgent Voluntary product Recall letters to customers on February 5, 2015 by e-mail with possible follow up on site by firm representatives. The letter identified the affected product, problem and actions to be taken. Customers are asked to quarantine inventory immediately and examine the affected lots for complete pouch seals. Any product found with an incomplete seal should be destroyed or returned to ZipLine Medical. Please complete the enclosed response form as soon as possible, even if you no longer have stock of the lot numbers subject to this correction. This information is required to reconcile the inventory. Please contact Customer Service at 1-888-326-8999 to coordinate any product returns, replacement stock, or credit notes.
Quantity
3109 - total - all model numbers
Distribution
Worldwide Distribution - US Nationwide and the countries of Panama, Tehran, Ireland, Sweden, Japan, Hong Kong, Finland, Pakistan, Taiwan, Malaysia.
Lot, serial or model codes
Lot 1002690, Exp 2015-08
510(k) numbers
not on file
PMA numbers
not on file
Product code
KGX
Firm FEI
3009673389
Firm city
Campbell
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register