FDA Medical Device Recalls (openFDA)
Ciba Vision Corporation: Quick Care Starting Solution. The product is used for cleaning and disinfection of soft (hydrophilic) contact lenses.
- Recall number
- Z-1234-03
- Recalling firm
- Ciba Vision Corporation
- Product
- Quick Care Starting Solution. The product is used for cleaning and disinfection of soft (hydrophilic) contact lenses.
- Status
- Terminated
- Initiated
- 2003-08-13
- Posted
- 2003-09-13
- Terminated
- 2005-01-28
- Reason
- Investigation has determined that the concentration of the isopropyl alcohol (IPA) decreased over time and may fall below specifications prior to reaching the expiration date.
- Root cause
- Other
- Action
- The consignees were notified of the recall by letter on August 13, 2003 asking for return of the product.
- Quantity
- 2,136 units
- Distribution
- CA, CT, FL, GA, IA, KS, MI, MN, OH, and WI
- Lot, serial or model codes
- Lot No. 1278080, exp. date 5/2004
- 510(k) numbers
- not on file
- PMA numbers
- ["P900029"]
- Product code
- LPN
- Firm FEI
- 1053053
- Firm city
- Duluth
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28