Roche Diagnostics Operations, Inc.: cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Recall number
Z-1233-2016
Recalling firm
Roche Diagnostics Operations, Inc.
Product
cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.
Status
Terminated
Initiated
2016-02-23
Posted
not on file
Terminated
2018-01-11
Reason
Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.
Root cause
Software design
Action
Roche Diagnostics sent an "Urgent Medical Device Correction" letter via UPS Ground (receipt required) on February 23, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To prevent this malfunction, a software patch has been developed and validated. The installation of this software patch is mandatory and will be performed by a Roche Field Engineering Specialist (FES). Your Roche FES will contact you to schedule the installation of the software patch. Until the software patch is implemented, Roche recommends that instrument operators monitor their systems for the occurrence of Error 140 Gripper has lost tube. If this error occurs, all components in contact with a spill must have extra cleaning. Refer to the cobas p 512 pre-analytical instrument Operators Manual Version 1.5. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 for further questions.
Quantity
35 Units
Distribution
US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA
Lot, serial or model codes
Part numbers:05083435001 and 06268854001
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register