Skip to the content

FDA Medical Device Recalls (openFDA)

Arjo, Inc.: BHM Medical Infrared (IR) Hand Control System, an optional wireless control component of the Arjo Maxi Sky 600 and BHM V4 Ceiling Patient Lifts; the hand control system is composed of two components, part 700.13640.33 - IR hand control, and part 492.00038 - IR receiver board

Recall number
Z-1233-2008
Recalling firm
Arjo, Inc.
Product
BHM Medical Infrared (IR) Hand Control System, an optional wireless control component of the Arjo Maxi Sky 600 and BHM V4 Ceiling Patient Lifts; the hand control system is composed of two components, part 700.13640.33 - IR hand control, and part 492.00038 - IR receiver board
Status
Terminated
Initiated
2008-02-20
Posted
2008-07-18
Terminated
2010-09-11
Reason
The ceiling lift may not stop lateral movement after releasing the left (or right) action button on a ceiling lift equipped with an infrared hand control.
Root cause
Device Design
Action
Arjo sent Urgent Device Field Correction Notification letters dated 2/15/08 on the BHM Medical letterhead to the end user accounts who received the affected IR remotes, advising them of the potential for the remote to become difficult to operate or to create uncontrolled movement of the ceiling lift. The accounts were instructed to remove the IR remote from service until the field correction is issued by BHM Medical. Instructions for removal of the IR equipment were provided in the Safety Advisory Notice included with the letter. The accounts were requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter, and indicating the number of units requiring correction.
Quantity
25 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
There are no serial numbers on the BHM Medical Infrared (IR) Hand Control System, part 700.13640.33 - IR hand control, and part 492.00038 - IR receiver board. The hand controls were installed on the following ceiling lifts: i) Maxi Sky 600, part LD10111, serial numbers LD-2605-15164, LD30733967, LD30733968, LD480627315, LD-5105-21404;  ii) V4, part 9100021, serial numbers V4-1107-29534, V4-1307-29837, V4-3307-33778, V4-1507-30147, V4-1905-14715, V4-4205-18404, V4-3606-26259, LD-1706-24108, V4-0206-22026, V4-3405-16157, V4-4606-27188, V4-4803-04671, V4-1307-29838, V4-1806-24149, V4-3307-33779, V4-4207-36144; part 9100021.13, serial numbers V4-0805-09557, V4-1006-23261; part 9100023, serial number V4-1406-23773; and part 9101021, serial number V4-0105-08044.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNG
Firm FEI
1419652
Firm city
Roselle
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register