Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corp.: Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.

Recall number
Z-1231-2023
Recalling firm
Datascope Corp.
Product
Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
Status
Open, Classified
Initiated
2023-02-07
Posted
2023-03-06
Terminated
not on file
Reason
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Root cause
Equipment maintenance
Action
Firm notified customers on February 7, 2023, in an Urgent Medical Device Correction letter which discussed four issues in total - this issue was Issue #4. Customers were instructed to contact their Datascope/Getinge service representative if they observe any visual damage to the O-ring installed as part of the quick disconnect fitting. Datascope/Getinge is currently updating the annual Preventive Maintenance instruction to include replacement of the quick disconnect fitting O-ring. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). On 11/04/2025 the recall was expanded, and consignees were notified via letter. The notification was for awareness to the update of Datascope s preventative maintenance in the IFU addendum, which was provided to consignees. The IFU Addendum contains an updated Schedule B Preventive Maintenance Schedule which calls for the Quick Disconnect Fitting O-Ring to be replaced and to lubricate the console release latch every 12 months or 2500 hours. The Service Manual has contained this update to Preventive Maintenance for the Helium O-Ring since June 2023. Consignees were instructed to review the IFU Addendum and local Preventive Maintenance Schedule and update their Preventive Maintenance schedule accordingly.
Quantity
11,792 units (5.408 US, 6,384 OUS)
Distribution
Nationwide US distribution. Worldwide international distribution.
Lot, serial or model codes
Model Number (UDI-DI): D998-00-0800-31 (10607567109053¿), D998-00-0800-32 (10607567111117), D998-00-0800-33 (10607567109008), D998-00-0800-35 (10607567109107), D998-00-0800-36 (10607567114187), D998-00-0800-45 (10607567108421), D998-00-0800-52 (10607567108438), D998-00-0800-53 (10607567108391), D998-00-0800-55 (10607567108414), D998-00-0800-65 (10607567113432), D998-00-0800-83 (10607567108407), D998-UC-0800-31 (10607567109053), D998-UC-0800-33 (10607567109008), D998-UC-0800-52 (10607567108438), D998-UC-0800-53 (10607567108391), D998-UC-0800-55 (10607567108414). All serial numbers
510(k) numbers
["K181122"]
PMA numbers
not on file
Product code
DSP
Firm FEI
3001418283
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register