FDA Medical Device Recalls (openFDA)
Datascope Corp.: Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
- Recall number
- Z-1231-2023
- Recalling firm
- Datascope Corp.
- Product
- Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
- Status
- Open, Classified
- Initiated
- 2023-02-07
- Posted
- 2023-03-06
- Terminated
- not on file
- Reason
- Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
- Root cause
- Equipment maintenance
- Action
- Firm notified customers on February 7, 2023, in an Urgent Medical Device Correction letter which discussed four issues in total - this issue was Issue #4. Customers were instructed to contact their Datascope/Getinge service representative if they observe any visual damage to the O-ring installed as part of the quick disconnect fitting. Datascope/Getinge is currently updating the annual Preventive Maintenance instruction to include replacement of the quick disconnect fitting O-ring. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). On 11/04/2025 the recall was expanded, and consignees were notified via letter. The notification was for awareness to the update of Datascope s preventative maintenance in the IFU addendum, which was provided to consignees. The IFU Addendum contains an updated Schedule B Preventive Maintenance Schedule which calls for the Quick Disconnect Fitting O-Ring to be replaced and to lubricate the console release latch every 12 months or 2500 hours. The Service Manual has contained this update to Preventive Maintenance for the Helium O-Ring since June 2023. Consignees were instructed to review the IFU Addendum and local Preventive Maintenance Schedule and update their Preventive Maintenance schedule accordingly.
- Quantity
- 11,792 units (5.408 US, 6,384 OUS)
- Distribution
- Nationwide US distribution. Worldwide international distribution.
- Lot, serial or model codes
- Model Number (UDI-DI): D998-00-0800-31 (10607567109053¿), D998-00-0800-32 (10607567111117), D998-00-0800-33 (10607567109008), D998-00-0800-35 (10607567109107), D998-00-0800-36 (10607567114187), D998-00-0800-45 (10607567108421), D998-00-0800-52 (10607567108438), D998-00-0800-53 (10607567108391), D998-00-0800-55 (10607567108414), D998-00-0800-65 (10607567113432), D998-00-0800-83 (10607567108407), D998-UC-0800-31 (10607567109053), D998-UC-0800-33 (10607567109008), D998-UC-0800-52 (10607567108438), D998-UC-0800-53 (10607567108391), D998-UC-0800-55 (10607567108414). All serial numbers
- 510(k) numbers
- ["K181122"]
- PMA numbers
- not on file
- Product code
- DSP
- Firm FEI
- 3001418283
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28