Skip to the content

FDA Medical Device Recalls (openFDA)

Quidel Corporation: Influenza Transport Swab Packages, Reference #20171, One lot with 1010 packages (each package contains 25 transport tubes). The Influenza Swabs in dry transport tubes are intended for use with influenza tests

Recall number
Z-1231-2017
Recalling firm
Quidel Corporation
Product
Influenza Transport Swab Packages, Reference #20171, One lot with 1010 packages (each package contains 25 transport tubes). The Influenza Swabs in dry transport tubes are intended for use with influenza tests
Status
Terminated
Initiated
2017-01-25
Posted
2017-02-22
Terminated
2017-06-02
Reason
Quidel Corporation is recalling Influenza Transport Swab Packages because it may include an incorrect sample swab mixed in with the proper swabs.
Root cause
Mixed-up of materials/components
Action
Quidel sent an Urgent Medical Device Recall letter dated 1/25/17 to all affected customers and distributors to inform them that Quidel is recalling the Influenza Transport Swab Packages because they may contain incorrect sample swab mixed in with the proper swabs. The letter informs the customers that the standard influenza swab is packaged individually, one (1) swab per package and uses a polyurethane foam tip. The incorrect swab is packaged as two (2) swabs per individual package and uses a rayon fiber tip. Customers are instructed to complete the enclosed Inventory Assessment/Certificate of Destruction form and return to Quidel. Contact Quidel Technical Support at technicalsupport@quidel.com or by calling 800.874.1517 (in the U.S.), or 858.552.1100 (outside the U.S.), for instructions on how to obtain replacement product. Our hours of operation are Monday through Friday, 7:00 a.m. to 5:00 p.m. Pacific Time.
Quantity
1,010 packages
Distribution
Nationwide Distribution to FL, GA, IL, IN, KY, MT, OH, PA, TX
Lot, serial or model codes
PS10171
510(k) numbers
["K141520"]
PMA numbers
not on file
Product code
OCC
Firm FEI
2024674
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register