Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiovascular Systems Inc: DIAMONDBACK 360 Peripheral 1.50 Classic, Model DBP-150CLASS145 Peripheral Orbital Atherectomy System (OAS) is a minimally invasive, catheter-based OAS designed for improving luminal diameter in patients with peripheral arterial disease (PAD). PAD is caused by the accumulation of plaque in the arteries of the leg or foot and reduces blood flow that may lead to pain, tissue loss, and eventual foot amputation, leg amputation or death. The system treats a broad range of plaque types in the lower limbs The DIAMONDBACK 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. The Stealth 360 Orbital PAD System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

Recall number
Z-1230-2017
Recalling firm
Cardiovascular Systems Inc
Product
DIAMONDBACK 360 Peripheral 1.50 Classic, Model DBP-150CLASS145 Peripheral Orbital Atherectomy System (OAS) is a minimally invasive, catheter-based OAS designed for improving luminal diameter in patients with peripheral arterial disease (PAD). PAD is caused by the accumulation of plaque in the arteries of the leg or foot and reduces blood flow that may lead to pain, tissue loss, and eventual foot amputation, leg amputation or death. The system treats a broad range of plaque types in the lower limbs The DIAMONDBACK 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. The Stealth 360 Orbital PAD System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
Status
Terminated
Initiated
2016-12-07
Posted
2017-02-22
Terminated
2017-03-23
Reason
CSI discovered that a 1.50mm Solid Crown OAD was labeled as a 1.50mm Crown OED.
Root cause
Employee error
Action
Cardiovascular Systems, Inc. sent an Urgent Medical Device Recall letter, dated 12/07/2016, to all affected customers. The letter identified affected product , stated the reason for recall, asked for product to be returned to CSI. Questions can be directed to Customer Service at 877-274-0901.
Quantity
8 units
Distribution
Nationwide Distribution to LA, MS, SC, TX
Lot, serial or model codes
Lot: 171686, Expiration: 2018-09-30
510(k) numbers
["K152694"]
PMA numbers
not on file
Product code
MCW
Firm FEI
3004742232
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register