Skip to the content

FDA Medical Device Recalls (openFDA)

Thoratec Corporation: Heart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own.

Recall number
Z-1230-2014
Recalling firm
Thoratec Corporation
Product
Heart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own.
Status
Terminated
Initiated
2014-03-04
Posted
2014-03-21
Terminated
2015-08-27
Reason
Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Root cause
Device Design
Action
Thoratec sent an Urgent Medical Device Correction letter dated March 4, 2014 to all affected customers. The letter was sent via e-mail. All distributors outside of the USA and Canada to whom affected product has been distributed have been notified of this field correction notice and have been provided copies of the Customer Letter to forward to their customers. The letter identified the affected product , problem and actions to be taken. Customers were requested to complete and sign the attached Acknowledgement Form and return it to Thoratec via fax (925-847-8571) or e-mail a scanned copy to customer.service@thoratec.com. For questions call 1-800-528-2577.
Quantity
7436 in total
Distribution
Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Lithuania, Poland, Saudi Arabia, Singapore, Slovenia, Sweden, Switzerland, The Netherlands, Turkey, and United Kingdom.
Lot, serial or model codes
Catalog number 105109 (found on side of unit) all serial numbers
510(k) numbers
not on file
PMA numbers
["P060040"]
Product code
DSQ
Firm FEI
2916596
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register