Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Bipolar cord for use with the VirtuoSaph Endoscopic Vein Harvesting System, MCBICORD1, Terumo Cardiovascular System, Ann Arbor, MI. The Terumo Bipolar cord MCBICORD1 is a flexible, reusable power cable that is indicated for use with the Terumo VirtuoSaph Endoscopic Vein Harvesting System and a compatible generator during saphenous vein harvesting procedures.

Recall number
Z-1230-2011
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Bipolar cord for use with the VirtuoSaph Endoscopic Vein Harvesting System, MCBICORD1, Terumo Cardiovascular System, Ann Arbor, MI. The Terumo Bipolar cord MCBICORD1 is a flexible, reusable power cable that is indicated for use with the Terumo VirtuoSaph Endoscopic Vein Harvesting System and a compatible generator during saphenous vein harvesting procedures.
Status
Terminated
Initiated
2011-01-10
Posted
2011-02-14
Terminated
2011-07-13
Reason
The Bipolar cord product specification indicates the product has a service life of 50 uses. However the IFU does not specify an end of service life. The firm has recieved 46 complaints on the Bipolar cord. The firm investigated and found that the units involved in the complaints were often over 12 months old. The firm believes the units involved in the complaints were beyond their product life
Root cause
Other
Action
Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE CORRECTION letter dated January 10, 2011, to Operating room managers. The letter informed the customers of the Bipolar cord product life of 50 uses. The letter states that the IFU for the bipolar cord is being updated to include the information regarding the limit of 50 uses and sterlization cycles. The letter recommends that user keep a sterile back up bi-polar cord in a convienent place during endoscopic vein harvesting procedures. The firm provided a tracking sheet in each package to help users track the number of sterlization cycles for the cord. Customers were instructed to review the Medical Device Correction notice, assure that all users were aware of the notice, and confirm receipt of the communicaiton by faxing the attached Customer Response Form to the fax number indicated on the form. For questions regarding this recall call Terumo CVS Customer Service at 1-800-521-2818.
Quantity
2782 cords
Distribution
Worldwide Distribution - United States including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, , NY, MO, MS, MT, NC, NE, NH, NJ, NM, OH, OK, OR, PA, SC, TN, TX, VA, WI, AND WV and the countries of Chile, Hong Kong, Malaysia, United Arab Emirates, Belgium, Japan, Korea, The Phillipines, Singapore, Taiwan, Thailand, Germany, and United Kingdom.
Lot, serial or model codes
01K, 02K, 03K, 04K, 05K, 06K, 07K, 08K, 09K, 0XK, 4ZK, 51K, 52K, 53K, 54K, 55K, 6XK, 6ZK, 71K, 72K, 73K, 74K, 75K, 76K, 77K, 78K, 79K, 7XK, 7YK, 7ZK, 81K, 82K, 83K, 84K, 85K, 86K, 87K, 87K, 88K, 89K, 8XK, 8YK, 8ZK, 91K, 95K, 96K, 97K, 99K, 9XK, 9YK and 9ZK.
510(k) numbers
["K083194"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register